Calibration
Flow Calibration
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
8-15
Figure 8–10.
Pull Forward on Handle of Filter Exchange
22.
Disconnect the Coarse flow tube from the front elbow fitting. These
are push connect fittings; you must pull back on the collet of the elbow
and then pull the hose with the barb to stem fitting out of the elbow.
23.
Leave the end of the hose open to the compartment air.
Figure 8–11.
End of Hose Open to the Compartment Air
24.
Align the roller assembly to the openings in the track and return the
holder into the track.
25.
Place the dual filter holder in the filter exchange mechanism and close
the filter exchange mechanism making sure the seals are not
compromised. Make sure you do not pinch the coarse flow tubing and
that it is still open to the compartment air.