Sampler Operation
Sampler Performance Verification
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
7-9
1.
Remove the flow audit adapter and install the Streamline Pro on the
sample tube (
Figure 7–7
). Other flow meters may require the use of the
flow audit adapter (valve open) to measure flow.
Figure 7–7.
Streamline Pro Installed on External Sample Tube
2.
Check the Fine, or high flow channel first. This is designated Flow on
the Sampler screen.
3.
Press
repeatedly to display the Main Menu. Scroll to Audit and
Calibration, press
, scroll to Audit, press
, scroll to Flow,
press
to turn on the flow and wait for the flow to stabilize. The
default sample flow should be 15 L/min, unless it was changed in the
default settings of the Sampler. Record the flow rate from the
Streamline Pro and the Sampler.
4.
Press
to turn Off the flow. The measured flow should be within
±5% of the displayed Current Flow. If this is not the case, perform the
flow rate calibration procedure.
5.
Check the Coarse, or low flow channel second. This is designated Flow
2 in the Sampler.
6.
Press
repeatedly to display the Main Menu. Scroll to Audit and
Calibration, press
, scroll to Audit, press
, scroll to Flow2,
press
to turn on the flow, and wait for the flow to stabilize. The
default sample flow should be 1.67 L/min, unless it was changed in the
default settings of the Sampler. Record the flow rate from the
Streamline Pro and the Sampler.