Sampler Settings
System Setup
6-6
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
The Standard Pressure (mmHg) is used by the Sampler to report the flow
rate results in standard terms. The default setting of Standard Pressure is
760. The default setting for Standard Pressure does not have any effect on
the volumetric flow rate and actual volume calculations by the unit. This
value is used to calculate standard sample volumes.
This parameter controls the filter fan. If this field is set to On, the filter fan
will run continuously. If this field is set to Auto, the filter fan will only
initiate if the temperature of the filter and ambient air differs by 2 °C. If
the Filter Fan field is set to Off, the filter fan will not initiate at all.
If no keys are pressed for 3 hours and the Auto Run feature is On, the unit
will automatically enter the Wait or Sampling mode and not wait for the
operator to press
.
The Site ID information has two 32-character fields. The user can enter
site identification numbers using one or both fields. If no entry is made in
these fields then no site identification numbers will appear in the Sampler’s
data output from this screen.
The Current date and time parameter are the current local date and time
(or other standard time selected by the user) expressed by default as hh:mm
and the current local date expressed by default as yyyy/mmm/dd.
Standard Pressure
Filter Fan
Auto Run
Site Identification
Current Date/Time