Calibration
Flow Calibration
8-10
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
g.
Use the menu key to escape back to the main menu
2.
Turn on the service mode from the main menu. Service mode is
enabled when the wrench icon is present in the status bar to the right of
the clock.
3.
From the main menu, go to Audit and Calibration and scroll to the
Calibration menu item.
4.
Before performing the flow calibration make sure the Ambient
Temperature and Pressure have been calibrated and are currently
indicating valid measurements that meet the audit criterion. Unit must
also be leak free, pass a leak test prior to calibration.
If you do not have the Flow Calibration Cassette, skip to “Using the Solid
Leak Check Disc:”.
Figure 8–1.
Flow Calibrationi Cassette
Using the Flow
Calibration Cassette: