Preventive Maintenance
WINS PM2.5 Impactor Maintenance
10-10
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
2.
Remove the WINS Impactor from its installed position in the Sampler.
The remove of the WINS Impactor is described in the Sampler
audit/verification section above.
3.
Separate the adapter from the WINS impactor (
Figure 10–4
).
Figure 10–4.
Adapter and WINS Impactor
Use the following procedure to clean the WINS impactor.
1.
Unscrew the two halves of the WINS impactor to separate its top piece
from its bottom section. This exposes the impactor assembly.
2.
Remove the impactor assembly from the bottom section of the WINS
impactor (
Figure 10–5
).
Figure 10–5.
Upper and Lower Sections of the Impactor Assembly (on left and
right sides of figure)
WINS Impactor
Cleaning