Installation
Standard Hardware Configurations
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
2-5
b.
Turn the dome connector on the bulkhead to ensure a tight grip.
c.
Place the PM
10
inlet on the end that has 3 cm (1.25 inches) of
machining. Ensure that the tube is pushed past the inlet’s two O-
rings (which provide some resistance) until it stops.
In its final sampling configuration, the entrance to the inlet must
be approximately 2 m (±0.2m) above the ground (6 to 7 feet). The
stand (57-004644) that comes with the Sampler provides a
convenient means of installing the Sampler to the correct height
requirements.
Figure 2–3.
Sample Tube with the PM
10
Inlet Installed
6.
Connect the unit to the electrical power source following all local
regulations. Be sure to fulfill all safety and regulatory requirements for
the hardware.
For 115 VAC configurations, the standard three-pronged U.S. plug is
provided at the end of the power cord. The unit is properly grounded
and use of a ground fault interrupter is not necessary.
For 220-240 VAC configurations, no electrical plug is provided at the
end of the power cord. This line must be wired in accordance with
local safety codes.
7.
Install the ambient temperature sensor.
a.
Locate the two screws on the left side of the enclosure.
b.
Remove the two screws. Be sure to retain the washers. This will
expose two holes.
c.
Locate the ambient temperature probe assembly in the compilation
package.