Calibration
Flow Calibration
8-18
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
Figure 8–16.
Filter Exchange Closed
37.
From the calibration screen scroll to Flow2 AND PRESS ENTER.
Begin the calibration. The unit should begin to bring the flow up to the
first set point.
ALLOW THE FLOW TO STABILIZE FOR AT
LEAST 15 MINUTES. THIS WILL ACT AS THE FLOW
CONTROLLER WARM-UP PERIOD.
After the first point please
wait 2-3 minutes for the flow to stabilize at the setpoint before entering
the measured flow.
38.
At the end of the calibration the unit will display the calculated
intercept or offset and span values necessary to linearize the flow sensor
to the operational range.
39.
Open the filter exchange mechanism and remove the dual filter holder.
40.
Pull the filter exchanger assembly forward aligning the rollers with the
opening in the track. Pull the bottom part of the holder out of the track
allowing you room to access the flow connections on the left side of the
filter holder bottom plate.
41.
Reconnect the Fine flow tube to the front elbow fitting. Make sure the
fitting is fully inerted into the elbow; you can feel it hit the stop.
42.
Calibration procedure is complete. Perform flow audit per the
operating manual instruction.