Firmware Screen Descriptions
Instrument Setup Menu
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
9-25
Temperature does not have any effect on the volumetric flow rate and
actual volume calculations by the unit.
●
In the Main Menu, choose Instrument Setup > System Setup >
Std
Temperature
.
SET STANDARD TEMPERATURE:
CURRENTLY: 25.0
o
C
SET TO: 25.
0
o
C
MOVE CURSOR
CHANGE VALUE SAVE
CTRST SSET INSTS STAT
The Average Pressure screen is used by the Sampler to maintain the proper
volumetric sample flow rate. The default value of Average Pressure is 999
mmHg, indicating that the unit should use the reading from the Sampler’s
ambient pressure sensor to maintain a constant volumetric flow rate.
●
In the Main Menu, choose Instrument Setup > System Setup >
Avg
Pressure
.
SET AVERAGE PRESSURE:
CURRENTLY: 999 mmHg
SET TO: 99
9
mmHg
MOVE CURSOR
CHANGE VALUE SAVE
CTRST SSET INSTS STAT
The Standard Pressure is used by the Sampler to report the flow rate results
in standard terms. The default setting of Standard Pressure is 999 mmHg.
The default setting for Standard Pressure does not have any effect on the
volumetric flow rate and actual volume calculations by the unit.
●
In the Main Menu, choose Instrument Setup > System Setup >
Std
Pressure
.
SET STANDARD PRESSURE:
CURRENTLY: 760 mmHg
SET TO: 76
0
mmHg
MOVE CURSOR
CHANGE VALUE SAVE
CTRST SSET INSTS STAT
Average Pressure
Standard Pressure