Calibration
Flow Calibration
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
8-17
32.
Pull the filter exchanger assembly forward aligning the rollers with the
opening in the track. Pull the bottom part of the holder out of the track
allowing you room to access the flow connections on the left side of the
filter holder bottom plate.
Figure 8–14.
Pull Forward on Handle of Filter Exchange
33.
Reconnect the Coarse flow tube to the front elbow fitting. Make sure
the fitting is fully inerted into the elbow, you can feel it hit the stop.
Disconnect the Fine flow hose from the back elbow fitting.These are
push connect fittings; you must pull back on the collet of the elbow
and then pull the hose with the barb to stem fitting out of the elbow.
34.
Leave the end of the hose open to the compartment air.
Figure 8–15.
End of Hose Open to the Compartment Air
35.
Align the roller assembly to the openings in the track and return the
holder into the track.
36.
Place the dual filter holder in the filter exchange mechanism and close
the filter exchange mechanism making sure the seals are not
compromised. Make sure you do not pinch the coarse flow tubing and
that it is still open to the compartment air.