Calibration
Ambient Air Temperature Calibration
8-2
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
1.
Press
repeatedly to display the Main Menu, scroll to Service
mode and press
. In the Service Mode screen, press
to
toggle the Service mode to On. Press
to return to the Main
Menu.
2.
Display the Audit and Calibration Menu by pressing
, scrolling
to Audit and Calibration, and pressing
to display the Audit and
Calibration menu.
3.
Enter the Calibration menu by pressing
, scrolling to Audit and
Calibration, and pressing
to display the Audit and Calibration
menu. With the cursor at Calibration, press
to display the
Calibration menu.
Use the following procedure to calibrate the ambient air temperature.
Note
The Service mode must be set to ON to perform the following
procedure. To turn the Service mode ON, refer to Step 1 in “
Calibration
Setup
” in this section.
▲
1.
If the calibration will be done at ambient conditions, the reference
temperature measurement can be taken with the probe installed in the
radiation sensor.
If the calibration is done at a temperature lower and/or higher than
ambient temperature, loosen the two screws on either side of the
temperature probe on the ambient temperature assembly and remove
the probe from the radiation shield. The reference and Sampler probe
should be banded together and immersed to the same depth in an
insulated constant temperature bath or block. Reference and Sampler
probes should be equilibrated at temperature for at least 5 minutes
before the temperature is measured.
In the Calibration menu, Scroll to Ambient Temp and press
.
2.
At the Ambient Temperature Calibration screen, use the pushbuttons
to enter the reference value in °C [°C = 5/9 x (°F - 32)], and press
to save the value. The Sampler automatically adjusts the
corresponding offset based on this input.
3.
If removed, reinstall the ambient temperature probe in the radiation
shield, and tighten the two screws on either side of the probe.
Ambient Air
Temperature
Calibration