Partisol 2000i Inlet Conversion
Converting a Partisol 2000i from a PM10 Sampler to a PM2.5 Sampler
4-8
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
1.
Open the Partisol 2000
i
enclosure door to access the filter exchange
mechanism.
2.
Pull the filter exchange mechanism toward you to open the filter
exchange mechanism.
3.
Lift the rollers through the slots in the left-hand and right-hand guides
(
Figure 4–7
).
4.
Allow the filter platform to settle downward, and remove the VSCC or
WINS impactor and adapter by pulling them straight downward.
5.
Remove the V-seal adapter from straight tube and install onto VSCC
or WINS impactor.
6.
Install the VSCC or WINS impactor with the adapter into the
enclosure by pushing them straight upward until they hit a stop.
7.
Install a filter carrier (with filter cassette) into the filter exchange
mechanism and close the mechanism by pushing the handle toward the
back of the enclosure.
The Partisol 2000
i
in now configured as a PM
2.5
Sampler (
Figure 4–13
).
There is no difference in Sampler programming or operation with the
Sampler configured as a PM
10
or PM
2.5
Sampler. The Sampler properly
measures and calculates sample volumes in Standard or Volumentric
conditions regardless of the Sampler configuration.