Firmware Screen Descriptions
Service Mode
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
9-89
OPTIONS:
>ENABLE I/O BOARD YES
CTRST SSET INSTS STAT
When an I/O expansion board is removed from the instrument, the Enable
I/O Board menu item should be set to “No” to prevent an I/O board alarm
from being triggered.
When the instrument includes an I/O board, the Enable I/O Board menu
item should be set to Yes so that an alarm will be generated if the board
fails.
At the Options menu, scroll to Enable I/O Board and press
to toggle
the function to Yes/No as appropriate.
The Display Pixel Test is used to test the LCD display. The display pixel
test screen is visible only when the instrument is in Service mode. For more
information on the Service mode, see “Service Mode” earlier in the chapter.
In the Main Menu, choose Service >
Display Pixel Test
.
DISPLAY PIXEL TEST:
DURING TEST PRESS
OR
x
TO EXIT,
TO TOGGLE
XXXXX
BEGIN TEST
GO BACK TO MENU
CTRST SSET INSTS STAT
The Service Mode screen is used to turn the Service mode On and Off.
●
In the Main Menu, choose
Service mode,
and press
to toggle the
Service mode On and Off.
SERVICE MODE:
CURRENTLY: OFF
SET TO: ON ?
TOGGLE VALUE
CTRST SSET INSTS STAT
Enable I/O Board
Display Pixel Test
Service Mode