Calibration
Flow Calibration
8-14
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
Continue with the following.
Figure 8–8.
Solid Leak Check Disk
19.
Scroll to Flow to perform the calibration of the 20 LPM or Fine
Sample channel.
20.
Open the filter exchanger and remove the dual transport holder. Place
the Solid Leak Check Cassette in the Coarse (Front) position of the
dual holder. Place a regular cassette with filter media in the Fine (Back)
position of the dual holder.
Figure 8–9.
Solid Leak Check Cassette in the Front Position
21.
Pull the filter exchanger assembly forward aligning the rollers with the
opening in the track. Pull the bottom part of the holder out of the track
allowing you room to access the flow connections on the left side of the
filter holder bottom plate.
Using the Solid Leak
Check Disc: