C-Link Protocol Commands
Communications Configuration
B-28
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
Table B–7.
Power Up Mode Values
Value
Power up Mode
0 Local/Unlocked
Mode
(default)
1 Remote/Locked
Mode
instrument id
This command reports the instrument id.
Send:
instrument
id
Receive:
instrument id 12
set instrument id
value
This command sets the instrument id to value, where value is a decimal
number between 0 and 127 inclusive.
Note
Sending this command via RS-232 or RS-485 will require the host to
use the new ID for subsequent commands.
▲
Send:
set
instrument
id
12
Receive:
set instrument id 12 ok
program no
This command the type and program version of the instrument.
Send:
program
no
Receive:
program no iSeries 2000iD 01.07.85.272
layout ack
This command reports the stale layout/layout change indicator (*) that is
attached to each response if the erec layout has changed since the last time
erec layout was requested, according to
Table B–8
. The following example
reports that the instrument is configured to do nothing.
Send:
layout
ack
Receive: layout
ack
0
set layout ack
value
This command disables the stale layout/layout change indicator (*) that is
attached to each response if the erec layout has changed since the last time
erec layout was requested, according to
Table B–8
.
Send:
set
layout
ack
Receive:
set layout ack ok