AK Protocol
Commands List
E-12
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
AK Protocol
Response if Command Addressed to Instrument is Unrecognizable
COM 2-WAY SETTINGS
RS-Para 1
52
ASCII code for the 1-digit Station Number (for example "4": 052). The Station Number is always 1 digit in length.
RS-Para 2
75048
ASCII code representation of the 2-digit Channel Number (for example: "K0": 075, 048). The Channel Number is
always 2 digits in length.
RS-Para 3
13010
Optional: Up to 3 ASCII codes can be added to response from instrument. In this case, <CR> and <LF> (ASCII
codes 013 and 010) are appended to the response. Enter 0 if nothing is to be appended.
RS-Para 4
0
Not used.
Transmission to Instrument
Response from Instrument
Byte
Example
Description
B
No Err
Error
Description
1
1
<STX>
ASCII CODE 002.
2
2
4
1-digit Station Number, RS-Para 1.
3
3
?
Question marks inserted in place of
unrecognized command.
4
4
?
Question marks inserted in place of
unrecognized command.
5
5
?
Question marks inserted in place of
unrecognized command.
6
6
?
Question marks inserted in place of
unrecognized command.
7
7
<space>
Space.
8
8
0
Number of current status conditions.
9
9
<space>
Space.
10
10
S
Syntax error.
11
11
E
Syntax error.
12
12
<ETX>
ASCII code 003.
13
13
<CR>
Up to 3 digits appended to the end
of the response transmission,
according to the entry for RS-Para 3.
14
14
<LF>
Up to 3 digits appended to the end
of the response transmission,
according to the entry for RS-Para 3.
15
15
16
16
17
17
18
18
19
19