Firmware Screen Descriptions
System Status
Thermo Fisher Scientific
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
9-63
Table 9–5.
Status Codes
Status
Codes
Description
Ok
No current status conditions.
M
Flash Memory. The Sampler detected an error in its flash memory.
Z
Power Failure. A power outage occurred during sampling on the current filter.
Power outage events of less than 60 seconds are not logged by the Sampler.
F1
Flow Out of Range. The measured sample flow rate through the flow channel
deviated by
5% from its set point for more than 5 minutes.
F2
Flow 2 Out of Range. The measured sample flow rate through the flow channel
deviated by
5% from its set point for more than 5 minutes.
S
User Pressed Stop. The user pressed the Run button while the unit was in the
WAIT or SAMP (sampling) mode.
S1
Flow Stop. The measured sample flow rate through the flow channel deviated
by
10% from its set point for more than 60 seconds. This is a critical status
condition, causing the Sampler to enter the Error or Wait Operating mode and
the status lights to blink.
S2
Flow 2 Stop. The measured sample flow rate through the flow channel deviated
by
10% from its set point for more than 60 seconds. This is a critical status
condition, causing the Sampler to enter the Error or Wait Operating mode and
the status lights to blink. (This status code applies to the 2000
i
-D only.)
A
Ambient temperature sensor. The ambient temperature sensor was not
installed correctly or indicated an invalid value. For ambient temperature, this
corresponds to a reading of less than -60
C or greater than 70
C.
T
Filter Temperature. One of the filter compartment temperature sensors was not
installed correctly or was out of range, i.e., less than -60
C or greater than 70
C.
PA
Ambient Pressure Sensor. The ambient pressure sensor is not installed
correctly or is indicating an invalid value (less than 300 mmHg or greater than
900 mmHg).
E
Electronics Temperature. The temperature of the unit’s electronics
compartment was outside of its usual operating range, i.e., less than 0
C or
greater than 70
C.
R1
Filter Temperature 1 Range. The measured temperature of Filter #1 (high flow
filter) exceeded the measured ambient temperature by more than 5° C for more
than 30 consecutive minutes.
R2
Filter Temperature 2 Range. The measured temperature of Filter #2 (low flow
filter) exceeded the measured ambient temperature by more than 5
C for more
than 30 consecutive minutes.
O1
Coefficient of Variation 1. The coefficient of variation of the sample flow #1
rate (high flow), expressed as a percentage, was greater than 2 (2%) during the
exposure of a filter.