KPM-35  Optical Power Meter User Manual

1

3

10

8

9

4

5

6

13

12

11

7

2

1

3

4

5

2

λ

850

nm

980

nm

1300

nm

1310

nm

1490

nm

1550

nm

1625

nm

REF/dB

dBm

REF

-5.00

270

HZ

OPM interface: insert the fiber to be tested, test the optical power.

1

LED lighting: flashlight lighting

2

Display: display test results and other information

3

λ key: can switch 7 test wavelengths

4

LED button: turn on/off the flashlight, press any button to turn 

on the backlight

5

REF/dB key: Short press the dB to switch unit, click once 

nW/dBm/dB to enter the upper clear data , press and hold 

until REF is displayed on the screen, and set the current optical 

power as reference value, enter the relative optical power test 

mode, the screen will display the setted reference .

6

On/off key: turn the device on or off

7

wavelength

9

Auto Off

8

Frequency

10

Battery level

11

Unit

13

Numerical value

12

External and key function description

Power on/Auto power off

     Short press the power button to turn 

on machine, and automatically start the 

auto-off function, the default auto-off 

time is 10 minutes. If you want to cancel 

this function, press and hold the power 

button .when starting up. When the 

power on icon disappears, it means to 

cancel the auto-off function.

Optical power reference value setting

Optical power wavelength switching

LED lighting

             Enter the optical power meter 

interface after booting, short press the 

"REF" key to set the current power value as 

the reference power, which can realize 

relative optical power test (insertion loss 

test) or absolute power test. After entering 

the relative power test mode, the insertion 

loss (dB) is displayed at the bottom of the 

screen, and the screen displays the 

reference value; short press the "dB" key to 

switch between linear power and absolute 

power display.

          The units of linear power, absolute 

power, and relative power are nW, dBm, 

and dB respectively.

P absolute power = 10lgP linear power / 

1mW.

P relative power = P absolute power-P 

reference power.

      Press the LED key to control the 

flashlight to turn on and off. Press 

any key to turn on the backlight.

         On the optical power meter 

interface, short press the "λ" key to 

switch the measurement wavelength. 

You can choose 7 different test 

wavelengths, 850nm/980nm/1300nm/

1310nm/1490nm/1550nm, 850nm, 

1625nm. To ensure the accuracy of the

test, the selected wavelength must be 

consistent with the wavelength of the 

measured optical signal