Firmware Screen Descriptions
System Status
9-64
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
Status
Codes
Description
O2
Coefficient of Variation 2. The coefficient of variation of the sample flow #2
rate (low flow), expressed as a percentage, was greater than 2 (2%) during the
exposure of a filter.
P
Elapsed Sample Period. For sampling programs of
12 hour duration without
conditional sampling, the Sampler issues this status code if the elapsed time of
a sample differed by
1 hour from the selected duration. Sampling programs of
less than 12 hours will not indicate an elapsed time error.
L
Leak Check Failed. This error code will appear if the leak check failed. A leak
check fails if the vacuum created in the sampling system leaks at a rate of 25
mmHg or greater, which is equal to the maximum leak rate of 80 mL/min that is
indicated as acceptable by the U.S. EPA.
D
Audit Performed. This status condition indicates to the user that an audit was
performed during sampling. It does not indicate an error. It is for information
purposes only. This status code appears only in the filter data screen.
B
Blank Filter. Indicates that “sample” was a zero time blank.
V
Hi Wait Flow. This status condition indicates there is flow greater than 1 L/min
through the system in Wait mode, when there should be no flow.
The Alarms menu displays a list of items that are monitored by the
Sampler. When an alarm is detected, the status of that associated item will
go from OK to FAIL. The number of alarms detected is displayed to
indicate how many alarms have occurred. If no alarms are detected, the
number zero is displayed.
The motherboard status, interface board status, and I/O expansion board
status (if installed) indicate that the power supplies are working and
connections are successful. There are no setting screens for these alarms.
●
In the Main Menu, choose System Status >
Alarms
.
ALARMS:
>ALARMS DETECTED 1
MOTHERBOARD STATUS OK
INTERFACE STATUS FAIL
I/O EXP STATUS OK
CTRST SSET INSTS STAT
Alarms