Firmware Screen Descriptions
Diagnostics
9-94
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
The Set Analog Outputs screen contains three choices: Set to full-scale, set
to zero, or reset to normal. Full-scale sets the analog outputs to the full-
scale voltage, zero sets the analog outputs to 0 volts, and reset returns the
output to normal operation. The analog outputs are returned to normal
operation upon exiting this screen. The following example shows the screen
when all analog outputs are set to “normal” operating mode.
●
In the Main Menu, choose Diagnostics
>
Test Analog Outputs >
ALL,
Voltage Channel 1–6
,
or
Current Channel 1–6
.
SET ANALOG OUTPUTS:
SETTING: ALL
OUTPUT SET TO: NORMAL
SET TO FULL SCALE
SET TO ZERO
RESET TO NORMAL
CTRST SSET INSTS STAT
The Contact Information screen displays the customer service information.
●
In the Main Menu, choose Diagnostics
> Contact Information
.
CONTACT INFORMATION:
CALL CENTER: 508-520-0430
WEB WWW.THERMO.COM
CTRST SSET INSTS STAT
Set Analog Outputs
Contact Information