Firmware Screen Descriptions
Audit and Calibration Menu
9-58
Partisol 2000
i
and Partisol 2000
i
-D Instruction Manual
Thermo Fisher Scientific
The Calibration for Flow menu is used to calibrate flow. The Calibration
for Flow2 menu items are used in the same manner as the Calibration for
Flow menu items.
●
In the Main Menu, choose Audit and Calibration >
Calibration >
Flow
.
CALIBRATION FOR FLOW:
>POINT 1
POINT 2
POINT 3
POINT 4
POINT 5
CTRST SSET INSTS STAT
SLOPE 1.00
INTERCEPT 0.00
CAL SETUP
The Setpoint for Flow screen is used to calibrate the flow.
●
In the Main Menu, choose Audit and Calibration >
Calibration >
Flow
>
Point 1-5
.
SP 1 FOR FLOW: 12.000
>FLOW MEASURED ACTUAL
STORED 12.00 12.000
LIVE 0.000 12.00
0
MOVE CURSOR
CHANGE VALUE SAVE
CTRST SSET INSTS STAT
The Slope menu item is a display only screen.
The Intercept menu item is a display only screen.
The Cal Setup Flow screen provides parameters for flow calibration setup.
●
In the Main Menu, choose Audit and Calibration >
Calibration >
Flow
>
Cal Setup
.
Flow
Setpoint 1-5 for Flow
Slope
Intercept
Cal Setup