background image

65 7120

Elektr. Digital

Oberflächen Rauheitsmessgerät 

Bedienungsanleitung

Ausstattung:

1. Ein/Aus-Schalter

2. LC Display

3. Funktions-Taste (mm/inch)

4. Funktions-Taste/Kalibrierung (Ra/z)

5. Schutzabdeckung

6. Messkopf

7. Batterie-Ladebuchse

8. Start-Taste

9. Reset-Taste

Spezifikation:

- Messungsparameter (µm):

Ra, Rz, Rq, Rt

- Prüfzuglänge (mm):

6

- Musterlänge (mm):

0,25 / 0,80 / 2,50

- Bewertungslänge (mm):

1,25 / 4.0

- Messbereich (µm):

Ra: 0,05 ~ 10,0

Rz: 0,1 ~ 50

- Anzeigefehler:

± 15%

- Abweichung der Anzeige:

<12 %

- Radius der Tastenspitze und Sensorwinkel:

10 µm ± 1 µm

Winkel: 90 +5° / -10°

- Messkraft der Tastspitzen:

≤ 0,

016N

Geschwindigkeit der Messkraft:

800 N/m

- Druck des Sensorkopfes:

0,5 N

- Batterie:

3,7 V, Lithium Batterie

- Abmessung:

106mm x 70mm x 24mm

- Gewicht:

200g

- Arbeitstemperatur:

-20°C ~ 40°C

- Luftfeuchtigkeit:

<90% RH

Lieferumfang:

-

Messgerät

-

USB-Ladekabel

-

Stecker

-

Kalibrier-Platte (Rauheitsnormal)

-

Bedienungsanleitung

-

Kunststoff-Tragekoffer

65 7120

Electronic Digital Surface

Roughness Measuring Device

Operation manual

Charge battery

Connect the charge cable to the roughness measuring device. Insert the plug in the respective socket (7) of the 
roughness measuring device. The charger then has to be connected with the mains or a computer. Charging 
duration is about 3 hours. Measurements can still be carried out during the charging procedure.

Device care

-

Protect the device against impacts and vibrations, excessive dust, moisture, oil 

stains and strong magnetic fields, etc.

-

If you are not using the device, please switch it off in good time in order to keep the energy consumption as 

low as

possible and charge within due time to avoid damage.
-

The sensor / measuring head is the precision unit of the testing device and should be treated with special 

caution. After each test, cover the measuring head carefully with the protected area in order to avoid undesired 
affects.

-

The enclosed standard plate for calibration should be carefully protected in order to avoid scratches that may

influence

the measuring accuracy.
-

In event of malfunctions, the user should under no circumstances open the device to attempt to make 

repairs on it.
Send the device back to the manufacturer along with the guarantee card and the relevant probe as well as 
a description so that it can be checked and repaired. To do so, please maintain contact with our customer 
representatives or the sales department of our company.

Troubleshooting:

Error

Possible cause

Troubleshooting

No display

Battery is flat

Charge the battery

Battery symbol 

appears in the 

display

Battery is flat

Charge the battery

Please note:

In order to exclude incorrect measurements, please perform the basic setting carefully.

Clean the measuring instrument on a regular basis using a cotton cloth, do not use benzine or 

solvent. Keep the measuring instrument dry as humidity may cause damage to it.

Electric etching or engraving may destroy the electronics, this shall void the guarantee!

Remove the battery if the device is not used for longer periods. Do not discard of used batteries in 

domestic waste! Dispose of used batteries at designated collecting points or as special refuse.