MEASUREMENT OPERATION
5-192
TMPM5200-2
Viscoelasticity measurement tab window
•
So
need not set it. The measurement pixels are fixed at
appear as follows:
urce signal
Since the source signal of the Viscoelasticity measurement tab window is set for
measuring viscoelasticity, you
256
×
256. When measurement starts, the images
Topography
(Topo Image)
Acos
φ
(Elasticity
Image)
Asin
φ
(Viscosity
Image)
You can confirm the scan mode and source-signal settings at viscoelasticity
measurement from the Viscoelasticity measurement (KFM) tab window by
moving directly to the advanced setting tab window.
Scan
mode:
4
Inputs
Measurement
mode:
VE-AFM
Source signal 1:
Topography
Source signal 2:
LIA1 (Aux1)
Source signal 3:
LIA2 (Aux2)
Source signal 4:
Topography
Although the scan mode is 4 Inputs during viscoelasticity measurement, the
source signal 4 is ignored, and the system actually acquires 3 images simulta-
neously.
•
Source Gain
The same as that in the AC-AFM tab window
•
Lock-in amplifier setting button
This button opens the Lock-in Amplifier Control window, which sets the extended
lock-in amplifier that is necessary for measuring viscoelasticity. Details on setting
the lock-in amplifier will be provided later.