MEASUREMENT OPERATION
5-188
TMPM5200-2
This button opens the I-V Curve setting dialog that sets the bias-voltage
change-width and number of accumul ons when you perform I-V Curve meas-
is button is operable only when you select the I-V Curve radio button or
Magnetic
•
I-V setting button
ati
urement. Th
the CITS radio button.
Force measurement tab window
•
orce measurement tab window is set for
t set it. The measurement pixels are fixed
Image
(Phase Image)
Source signal
Since the source signal of the Magnetic F
measuring the magnetic force, you need no
at 256
×
256. When measurement starts, the images appear as follows:
Topo
Magnetic
Image
You can confirm the scan mode and source-signal settings at magnetic force
measurement from the Magnetic Force measurement tab window by moving
directly to the advanced setting tab window.
Scan mode:
4 Inputs (at P_Lift) or 2 Inputs (at L_Lift)
Measurement mode:
P_Lift or L_Lift
Source signal 1:
Phase
Source signal 2:
Topography
Source signal 3:
Topography
Source signal 4:
Phase
While you use the Point by Point lift mode, the scan mode becomes 4 Inputs
and the system displays the topographic image as the average of source signals
2 and 3, and the magnetic force image as the average of source signals 1 and 4.