4 PREPARATION FOR OBSERVATION
4.1.3 Mounting the Scanner Holder
1.
Turn off the SPM CONTROL unit.
Be sure to turn off the power to the SPM CONTROL unit in advance
when mounting or dismounting the scanner holder, specimen
holder or SPM head unit. There is a risk of electric shock due to the
high voltage applied to the connector pins from the SPM CONTROL
unit.
WARNING
2.
Select the scanner holder having the most suitable scanning range according
to the specimen to be observed (if you have optional scanner holders). Insert
it into the specimen stage along the keyway. Then, lock it with the two
fastening screws.
Engraving
Keyway
Specimen stage
Fastening screw
Scanner holder
When you replace the scanner holder with an optional one, you have to designate the
new one using the WinSPM software. For the designating procedure.
Refer to Section 5.10.1, "SPM Scan Menu."
The engravings of the scanner holder are as follows.
Scanning range
Engraving
Type of scanner
X Y
Z
STD Standard
scanner
10
3
NSC
Narrow area scanner
1
1
WSC
Wide area scanner
80
5
WZS
Scanner for biological material
50
10
SWZ
Scanner for rugged surfaces
30
15
4-4