5
MEASUREMENT OPERATION
5-250
TMPM5200-2
e Total Size input boxes. The area specified in
these input boxes shows the total size of scanning.
The image will appear as below if the parameters are as follows:
•
Number of Images
2
×
2
Total Size
Specify the total scanning area in nm in th
•
Total Size
X: 3000 nm,
Y: 3000 nm
•
Overlap
X: 0 nm,
Y: 0 nm
1500
1500
3000
he scanning area of an image will appear in the Image Size display position of the
dow.
3000
T
Montage win
If you select Overlap specifying X: 10 nm and Y: 10 nm, the image will appear as below.
1505
1505
3000
3000
Y 10 nm
Overlap X 10 nm
During montage scan, Scan–Size in the SPM Parameters window is not in effect.
In montage scan, this Total Size has priority in determining the scanning range of one
frame.