4 PREPARATION FOR OBSERVATION
Specimen stub
Scanner holder
Specimen stub
installing rod
4.1.9 Specimen
Exchange
When you replace the specimen just observed with a new specimen, you do not need to
turn off the SPM CONTROL unit.
Never exchange a specimen while a high bias voltage is being applied to
the specimen. You might get an electric shock due to the high voltage
applied to the specimen.
WARNING
1.
Click on SPM Scan on the main menu of the WinSPM software and select
Scan
…
from the pull-down menu. Then, select the Advanced tab. The SPM
Parameters window will be displayed.
2.
Select the Tip button, and the Tip window will be displayed.
3.
Confirm that the High Bias check box
□
located at the bottom left of the Tip
window has not been checked.
If the check box has been checked
click on it to remove the check mark from the
check box.
TMPM5200-2