TMPM5200-2
5-1
5.1 BASIC USAGE OF WinSPM
5.1.1 Before
use
Before using WinSPM, confirm that the system is properly set up and that the calibrated
values of scanner sensitivity and motor drive position have been recorded. If and when it
becomes necessary to re-install the WinSPM software, you must save these values in
WinSPM once again.
For details, refer to Sect. 7.2, “Setting Up the WinSPM Program.”
5.1.2 Glossary
Basic terms used in discussing WinSPM are explained below.
Administrator’s privilege
In WinSPM, the name of a user (normally a system administrator) who has access
privileges to windows relating to system maintenance is registered as “Admin”. Once the
name of the administrator has been registered, no other name can be registered
as ”Admin”.
If you log into WinSPM as an “Admin”, you can perform the following operations.
•
Setting, altering or saving the calibrated values of the scanner
•
Starting special measurement mode to calculate the calibrated values of the scanner
•
Altering the power supply voltage
•
Setting, altering or saving the calibrated values of the motor drive position
Parameters such as measurement and display conditions saved with the (Admin)
user name compose the basic parameters.
The password for user name “Admin” is saved by default as JEOLSPM (with all
characters in uppercase). Before using the instrument, be sure to change the pass-
word. We recommend that a password for the person responsible for the SPM
system be used.
The Administrator’s privileges involve changing instrument conditions such as the
power supply voltages and other important settings. If they were to be changed
erroneously, malfunction of the instrument might occur. To avoid this, start up
WinSPM with the user privilege described below.
User’s privilege
Users’ names other than “Admin” are designated for enabling all measurement operations
except those relating to maintenance. Use these privileges in routine operation.