ver into an approach state
If the lock-in amplifier meter needle swings out of scale (or in the case of
Model 5110, the OVLD LED lights up), raise its sensitivity using the Lock-In
Amplifier window.
9.
Set oscilloscope (optional) CH2 to AUX1 to monitor CH2; then reset Filter
and Gain in the KFM window according to the potential resolution.
8.
Click on the Retract Off button and set the cantile
again.
At this stage, feedback will be executed to set the
ω
component (one cycle) to
zero.
TMPM5200-2