MEASUREMENT OPERATION
■
Lithography
This opens the Lithography dialog that you should set for applying this instrument to
lithography.
Settings
Specify the parameters for lithography in this frame of the Lithography window:
Parameters
Description
Preamp Gain
Specify the gain for preamplifier in this selection box.
Current/nA (for STM mode)
Specify the tunneling current in this input box.
Reference/V (for AFM mode)
Specify the reference voltage in this input box.
Sample Bias/V
Specify the bias voltage to the sample in this input box.
Toggle Sample/Hold
Select whether or not the feedback circuit should be turned
on by checking this box.
Add External X, Y, or Z Signal
Select whether or not the external signal should be added by
checking this box.
Time Period/ms
Specify the duration for applying Current/nA or Reference/V
in this input box. If 0.0 is specified, Current/nA or
Reference/V is applied while pressing the left mouse button.
This function is in effect only for Trace mode.