Image to analyze (image on the left)
The image for particle analysis and its SPM information are displayed.
Binary digitized image (image on the right)
Image to analyze digitized in a binary value. The threshold level is displayed at the left
of the screen. The threshold level to digitize in a binary value is set from the particle
analysis setting window.
Particle information (bottom of binary digitized value image)
Here is displayed the information about particles extracted by binary digitized value
processing.
•
Mean area
Mean area of groups of particles extracted
•
Mean perimeter
Mean perimeter of groups of particles extracted
•
Total area
Sum of all areas of groups of particles extracted
•
S (area) ratio
Ratio of total area of groups of particles to the total area in the
image
•
No. of particles
Total number of particles extracted
Information can be displayed by setting a threshold value from the particle analyze
setting window and executing analysis.
Statistic information
The statistic processing result is displayed as histogram by setting the threshold value
and executing analysis from the particle analyze setting window.
The statistic specie to display is set from the particle analyze setting window.
■
Particle Analysis Setting Window
Setting tab window
Tab window that performs various settings and analysis processing, such as setting of
threshold value level for binary value digitizing processing, processing of extracted area
and setting of decision system.
•
Threshold value setting
TMPM5200-2
6-57