MEASUREMENT OPERATION
5-142
TMPM5200-2
4.
When the cantilever has moved 0.1 to 0.2 mm away from the specimen, click
on the Stop button to stop the movement.
5.
Turn the manual approach screw to move the specimen stage down.
Stage moves upward
SPM head unit
Stage moves downward
Groove
Manual approach screw
Cantilever holder
Tip holder
The figure above shows the JSPM-5200.