6
ANALYSIS OPERATION
Normalization Parameters
Low conductance: Value that appears black on the display. This parameter is used only
in the Conductance display mode.
High Conductance: Value that appears white on the display. This parameter is used only
in the Conductance display mode.
Images to Display
•
Show Topographic Image
When this check-box is checked, a topography image is displayed in the top left
frame of the Display Window.
•
Images to Select
The current image series to be displayed is selected in this selection box.
None:
Nothing is displayed.
Spaced:
Images are displayed in equal voltage intervals.
First:
Designated number of images starting with the first voltage value are
displayed. Number is designated in “Number to display.”
Last:
Designated number of images starting with the last voltage value are
displayed. Number is designated in Number to display.
Mixed:
Images for designated voltages are displayed.
6.11.2 Special Displays
Since one CITS consists of a lot of data, special display modes are provided to display
CITS images.
Display Mode
Description
Paged display
64
×
64 display
Rapid display
Sixteen frames, beginning with the first one, are displayed.
Sixty-four frames are displayed at a time with a size of 64
×
64 pixels for
each frame. If the number of images is 16 or less, these images are
displayed at a size of 128
×
128 pixels. When the “+” cursor is moved to
any of the sixty-four image frames, the information of the image is
displayed on the “Data Information” window.
All the CITS images are sequentially displayed in real time as a moving
picture on the top left 128
×
128-pixel frame with scroll bar operation.
Memor y
Paged display (64
×
64 display)
CITS data for
one frame
64 frames
Left
mouse-button
Rapid display
64 frames
CITS data for
one frame
Displayed in the
top left corner
Bar-char t
operation
Displays
64 frames
6-94
TMPM5200-2