Condition in which the specimen and cantilever tip are completely separated
(condition prior to approach)
Neutral position (condition at the end of approach)
Condition in which the cantilever tip is in contact with the specimen
■
Terminating observation
Terminate observation according to the following procedure:
1.
Stop scanning.
2.
Click on the Retract On button to retract the cantilever tip.
3.
Click on the Out radio button for Z and click on the Start button.
4
e specimen, click
5
ge down.
.
When the cantilever has moved 0.1 to 0.2 mm away from th
on the Stop button to stop the movement.
.
Turn the manual approach screw to move the specimen sta
5-34
TMPM5200-2