5
MEASUREMENT OPERATION
TMPM5200-2
5-219
■
Advanced Control
Whe
The
other
The
parameters.
Scan tab window
Specify detailed parameters that are necessary for a scan.
n you click on the Advanced Control button, the Advanced Control window opens.
Advanced Control window includes some parameters that are not often used, and
parameters that are necessary to perform optional measurements.
Advanced Control window has three tab windows in which you can specify detailed
•
External S
This controls ON or OFF for the signals that enter the SPM CONTROL unit.
•
Add External X Signal
When this check box is clicked on, the signal input from the ADX terminal on
the rear of the SPM CONTROL unit is added to the X-axis scanning signal.
The signal input from the ADX terminal is increased to a signal 20 times
greater in voltage through the piezo-d
That is, input-
the power-supply
•
Add External
When this check box is clicked on, the signal input from the ADY terminal on
the rear of the SPM CONTROL unit is added to the Y-axis scanning signal.
The signal input from the ADY terminal is increased to a signal 20 times
greater in voltage through the piezo-drive high-power amplifier. That is, input-
ting a
±
1 V sine wave to the ADY terminal outputs
±
20 V.
The maximum input rating of the ADY terminal is
±
10 V. However, if the sum
of the input voltage and Y-axis scanning voltage exceeds the power-supply
ignal
rive high-power amplifier.
ting a
±
1 V sine wave to the ADX terminal outputs
±
20 V.
The maximum input rating of the ADX terminal is
±
10 V. However, if the sum
of the input voltage and X-axis scanning voltage exceeds
voltage of the SPM CONTROL unit, the maximum X-axis scanning voltage is
saturated at the power-supply voltage.
Y Signal