TMPM5200-2
5-93
due to the initializa-
5.
Spec
It will take about one second until the Amplifier starts up
tion of parameters. If errors should happen, confirm connections and settings of
cables.
ify the parameters according to the following table.
Sensitivity
1 mV
OSC Frequency
20.000
Time Constant
1 ms
OSC Level
0.200
Ref. Phase
0.000
Output
ACosT, ASinT
6.
Click on the Advanced button.
The Lock-In-Amplifier 1 Advanced window opens.
7.
Specify the parameters according to the following table.
Filter
Band pass
Dynamic Res.
High Resolution
Filter Frequency/kHz
20.000
Offset
□
Line Reject
Off
Track
Ref
INT
2F
□
Slope/12dB
□
8.
Click on the Retract Off button in the Contact Mode tab window.
9.
Click on the Auto Configure button in the Lock-In-Amplifier window.
The lock-in amplifier will adjust the phase and gain automatically.