5-120
TMPM5200-2
The present cantilever position is indicated by the “
+
” mark. Usually, the initial
position is the top-left corner of the frame.
Current
image
Topography
image
Red line
Cantilever position
Cursor
Green line
in which the image is
a red line is not active now. If
will
5.
Move the cursor to the cantilever position (“+”); then drag the “+” mark, while
holding down the left mouse button, to the point where you want to perform
I-V m
The cursor is in effect only inside of the active window
surrounded by a green line. The image surrounded by
you acquire the current image according to “Current Image Observation”, the image
be grabbed later and becomes an active window.
easurement.
Topography
Measurement
position
The cantilever is actually moving in real time during the cursor dragging, so be sure
to drag the “
+
” mark slowly.
6.
After placing the “+” mark at the measurement point, press the right mouse
button to finalize the position.
The Tip window will be displayed.
7.
Deselect Tip Retract.
8.
Click on the SPS button.
The SPS Parameters window will appear.