4 PREPARATION FOR OBSERVATION
4.1.7 Inserting the Cantilever Holder or the Tip Holder into the
SPM Head Unit
1.
In order to prevent the cantilever or the tip from colliding with the scanner
holder, confirm that the specimen stage is lowered to a low enough position.
If it is not, move the specimen stage downward by turning the manual
approach thumbscrew.
SPM head unit
Specimen stage moves upward (UP)
Manual approach thumbscrew
Groove
Specimen stage moves downward (DOWN)
Tip holder
Cantilever holder
2.
Insert the cantilever holder or the tip holder into the SPM head unit groove
until it touches the innermost wall.
TMPM5200-2
4-7