5
MEASUREMENT OPERATION
5-262
TMPM5200-2
resistance or other resistances. This causes the sensor to display a higher temperature on
the dialog than the actual temperature. Set an offset to correct this error.
When the actual temperature of the specimen is 20
℃
(the temperature is measured
before heating or cooling), if the indicated temperature on the dialog under Offset 0
appears 25
℃
, set –5 for Offset.
The temperature offset applies for both display temperature and setting temperature.