TMPM5200-2
5-117
•
Affixing a specimen for JSPM-5200/JSPM-5700
In order to observe a current image correctly, it is essential that good conductivity
be kept between the specimen and the specimen stub. Affix the specimen to the
specimen stub with conductive tape or conductive paste such as Conductive paste.
If possible, apply the conductive paste not only to the surface but to the side of the
specimen as well.
Specimen
Conductive paste
Specimen stub
figure above shows the specimen for the JSPM-5200.
The
Selecting a cantilever
d short, as shown in the figure below and you
•
If there are two cantilevers, long an
want to use the short cantilever for image observation, be sure to remove the long
one using a pair of tweezers.
Long
cantilever
Tweezers
Laser
Cantilever
The top of the long cantilever
use a short cantilever out of two cantilevers,
long and short, be sure to remove the long one with a pair of
e top of the long cantilever touches the
touches the specimen before
the shor t one.
—— CAUTION ———————
When you select and
tweezers. Otherwise, th
specimen before the short one does.