7
APPENDIX
TMPM5200-2
7-15
7.3 CANTILEVER
The cantilever is a consumable item. Select an appropriate cantilever according to the
mode and purpose of the measurement.
The following are the specifications for the cantilevers manufactured by Olympus,
Nanosensor, and MDT. Please consult your JEOL service office for purchasing.
Olympus
Mode
Model
Lever
shape
Length
(
µ
m)
Width
(
µ
m)
Thick
ness
(
µ
m)
Tip shape
Tip
length
(
µ
m)
Spring
constant
(N/m)
Resonance
frequency
(kHz)
Metal
coat
Tip
diameter
Number
of tips
OMCL-AC120TS-1
Strap 120 30
4
Tetrahedral
10 57 300 Single
side
< 10
15
OMCL-AC120TN-1
Strap 120 30
4
Tetrahedral
10 57 300 None
< 10
15
AC mode
OMCL-AC240TS-1
Strap 240 30
4
Tetrahedral
10 7.2
65 Single
side
< 10
15
OMCL-TR400PS-1
Triangle 100
200
―
0.4
Pyramidal
2.9 0.09
0.02
40
13
Single
side
< 20
35
OMCL-TR400PS-2
Triangle 100
200
―
0.4
Pyramidal
2.9 0.09
0.02
40
13
Single
side
< 20
35
OMCL-TR800PS-1
Triangle 100
200
―
0.8
Pyramidal
2.9 0.68
0.16
86
27
Single
side
< 20
35
OMCL-TR800PS-2
Triangle 100
200
―
0.8
Pyramidal
2.9 0.68
0.16
86
27
Single
side
< 20
35
OMCL-TR400PB-1
Triangle 100
200
―
0.4
Pyramidal
2.9 0.10
0.025
36
11
Double
side
< 20
35
Contact
mode
OMCL-TR800PB-1
Triangle 100
200
―
0.8
Pyramidal
2.9 0.77
0.18
78
25
Double
side
< 20
35
OMCL-RC800PS-1
Strap 100
200
20
40
0.8
Pyramidal
2.9 0.37,0.05
0.75,0.09
88
22
Single
side
< 20
35
FFM
OMCL-RC800PB-1
Strap 100
200
20
40
0.8
Pyramidal
2.9 0.37,0.05
0.75,0.09
88
22
Double
side
< 20
35
* The above specifications are nominal and are subject to change. The values of the spring constant and resonance
frequency were obtained by simulated.