5-100
TMPM5200-2
■
Setting
When u
1.
Click on the Lock In button.
The Lock-In Amplifier dialog box opens.
the Lock-in Amplifier
sing Model 7265
2.
S
ors should happen, check the power supply, connections
3.
elect Connect to connect the lock-in amplifier.
It will take several seconds until the amplifier starts up due to the initialization
of parameters. If err
and settings of cables.
Specify the parameters according to the following table.
Sensitivity
50 mV
Time Constant
5 ms
Ref. Phase
0.000
Ref. Frequency
3F
Output
ACos
・
ASin
OSC Frequency
20.000
OSC Level
5.0
Use result (1F freq)
Select
Start
Frequency/kHz
60 kHz
Stop
Frequency/kHz
80 kHz
4.
Click on the Retract Off button of the Contact Mode tab window.
5.
Click on Start in Modulation Signal Scan.
Scan starts. The peak is being automatically searched and the oscillation frequency
will be set. If the peak position is displayed in reverse, click on the +90 button and
start scan again.
6.
Set Ref. Frequency to 3F.
7.
Click on Auto Phase to adjust phase.