INSTRUCTIONS

PINHOLE UNIT

A X 7 3 8 3

IX2-RFSPOT

The IX2-RFSPOT has been designed to be used in place of the field iris diaphragm (FS) of a IX2 Series reflected
fluorescence illuminator (IX2-RFAL). When the IX2-RFSPOT is used, a desired pinhole

*

 can be projected on the

specimen plane, which is convenient for canceling the cage on caged specimens.
 

*

No pinhole is provided with the unit. Please fabricate a pinhole yourself (see the figure illustrated later in this
manual) or use the commercially available recommended product (pinhole manufactured by Melles Griot, outer
diameter 16 mm).

1

    External View

2

    Precautions for Operation

 · Flare is produced around the pinhole. Ghost may be noticeable when the pinhole position is deviated from the center of

the field of view.

 · The pinhole image 

may not be able to be focused perfectly depending on the objective and wavelength in use.

 · Centering of the pinhole guarantees only the movement in the center of the field of view. Movement to the peripheral

section of the field is not possible.

 · The pinhole position may be deviated slightly when the pinhole is engaged or disengaged. Perform centering whenever

strict positioning is required.

 · The pinhole cannot be set to an intermediate position between the IN and OUT positions.

 · The IX2-SFR stage gets in the way of centering the pinhole. To prevent it, remount the IX2-SFR with the front in the back or

replace it with the IX-SVL-2.

IX2-RFSPOT

Lamp housing side

Centering screws

Pinhole engagement knob

Pushed in: Pinhole engaged (IN)

 Pulled out: Vacant hole (OUT)

Clamping screw

Focus adjustment screw

Microscope frame side

Washers

(x 4)

Clamping ring

Summary of Contents for IX2-RFSPOT

Page 1: ...is deviated from the center of the field of view The pinhole image may not be able to be focused perfectly depending on the objective and wavelength in use Centering of the pinhole guarantees only the movement in the center of the field of view Movement to the peripheral section of the field is not possible The pinhole position may be deviated slightly when the pinhole is engaged or disengaged Per...

Page 2: ...ow the instructions given with the figure if the pinhole is fabricated by you or insert with the indications facing down the lamp housing side if the pinhole is a commercially available product Attaching the Pinhole Unit Insert the pinhole unit in the microscope position where the field iris diaphragm has been and clamp the pinhole unit using the field iris diaphragm clamping screw 4 Operation Per...

Page 3: ...n Unit mm Pinhole position Microscope frame side Size indication surface of common commercially available pinhole Lamp housing side 1 6 1 1 16 0 1 0 3 UIS2 Series Objective Type Magnification Excitation Wavelength Priority Brightness Contrast 340 nm 360 nm 480 nm Brightness Contrast Brightness Contrast Brightness Contrast Number of Washers Number of Washers Number of Washers UPlanFLN40X X X X X 1 ...

Page 4: ...st position where the pinhole appears sharpest although it looks a little dim by adjustment of the mercury lamp s collector lens Usable regardless of the model number 3 2 or none Number of Washers Number of Washers Number of Washers UApo20XW3 340 0 0 0 0 2 2 UApo20X3 340 0 0 1 1 2 2 UApo40XOI3 340 0 0 0 0 2 2 UApo40XW3 340 0 0 0 0 2 2 UPlanApo40X X X 0 0 2 2 UPlanApo40XOI3 X X X X 2 2 PlanApo40XWL...