High Technology, Inc.
109 Production Road,
Walpole MA 02081 USA
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared by
J.S. Bolio
Approved by
S. Titov
Effective Date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 221 of 222
(0756) ISE Module - Air in
Cleaner
Insufficient cleaning solution
pipette into ISE module sample
entry port.
Check cleaning solution tube
Contact Technical Support.
Waste peristaltic pump failure.
Contact Technical Support.
Fibrin or salt is plugging the
electrode flow path.
Use cleaning procedure.
Remove electrodes and clean or
replace electrode with plugged flow
path. Reinstall electrodes and
recalibrate.
(0757) ISE Module - Air in
Segment
Sample cup contamination. The
sample cup may become dirty (i.e.
possible protein buildup, salt from
reagent, dust falling in from
above) during normal use.
A cotton tip applicator dipped in DIH20
to clean around and in the sample cup,
being careful not to dislodge any
particles that may obstruct the flow path.
Air bubble in sample
– if the
sample probe delivering the
sample or any reagent is to fast or
forceful air bubbles can be
created and drawn into the flow
path.
Contact Technical Support.
(0758) ISE Module - Pump
Calibration error
Insufficient saline solution pipette
into ISE module sample entry
port.
Check saline solution tube
Contact Technical Support.
Waste peristaltic pump failure.
Contact Technical Support.
Fibrin or salt is plugging the
electrode flow path.
Use cleaning procedure.
Remove electrodes and clean or
replace electrode with plugged flow
path. Reinstall electrodes and
recalibrate.
(0759) ISE Module - No Flow
Pump tubing obstructed
Contact Technical Support.
Peristaltic pump failure
Contact Technical Support.
(0760) ISE Module - Bubble
Detector
Fluid leaks
Contact Technical Support.
Pump tubing obstructed
Contact Technical Support.
(0761) ISE Module -
Reagents Kit Electronic Key
Read
Internal error.
Contact Technical Support.
(0762) ISE Module -
Reagents Kit Electronic Key
Write
Internal error.
Contact Technical Support.
(0763) ISE Module - Invalid
Command
Internal error.
Contact Technical Support.