High Technology, Inc.
109 Production Road,
Walpole MA USA 02081
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared By
J.S. Bolio
Approved By
S. Titov
Effective date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
Page 16 of 217
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Dispensing Arm and Hydraulic System
Samples and reagents preheated to 37ºC. High precision diluter
Maximum aspiration volume
: 500 µl (physical capacity of the diluter)
Internal and external washing of the probe through liquid diaphragm pump Volume detection
Probe impact sensors
Cuvette washer
Water consumption
: 2.4 ml/cuvette
Uninterrupted washing and drying of cuvettes as required by dispensing or maintenance process.
Photometer
Interferential filters
: 340, 380, 405, 450, 505, 546, 578, 600, 650 and 700 nm and an extra filter as
optional
Krypton-halogen lamp
: 12 volts 20 watts
Optical path
: 6 mm
Photometric range
: -0.100 to 3.600 A.
Type of measurement
: monochromatic and bichromatic
ISE module (optional)
Electrodes
: four ion-selective electrodes Li+, Na+, K+, Cl- and one Reference electrode
Peristaltic
pumps
: three
Reagent
Kit
: Calibrant A, Calibrant B, Waste storage.