Operator’s Manual: HTI BioChem FC-360
Page 216 of 217
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Electrical noise spike from
environmental source.
Contact Technical Support.
Component failure on ISE Module
board.
Contact Technical Support
(0705) Drift in Serum/Plasma
If single electrode is flagged: May
occur when new electrode or
Calibrant A is installed.
Purge the Calibrant A and recalibrate
the ISE module. If the electrode is new
it may initially drift as it rehydrates over
the course of 15 minutes.
If single electrode is flagged:
Replace the electrode and recalibrate.
If multiple electrode are flagged:
May occur when new electrode or
Reagent Kit is installed.
Purge the Calibrant A and recalibrate
the ISE module.
If multiple electrode are flagged:
Reference electrode.
Replace reference electrode and
recalibrate.
Electrical noise spike from
environmental source.
Contact Technical Support.
Component failure on ISE Module
board.
Contact Technical Support
(0706) Out of range in
Serum/Plasma
Samples results are out of range
limits:
Whole blood, serum, plasma
Li+
0.20
– 3.50 mmol/L
Na+ 100.0
– 200.0 mmol/L
K+
1.00
– 10.00 mmol/L
Cl-
50.0
– 150.0 mmol/L
Check quality of sample.
Run Calibration procedure, run control
samples and retest.
(0707) mV Out in Urine
Salt is plugging the electrode flow
path.
Remove electrodes and clean or
replace electrode with plugged flow
path. Reinstall electrodes and
recalibrate.
If single electrode is flagged
Replace problem electrode and
recalibrate.
If multiple electrodes are flagged
Replace reference electrode and
recalibrate.
Electrical noise spike from
environmental source.
Contact Technical Support.
(0708) mV Out in Calibrant B
If single electrode is flagged: May
occur when new electrode or
Calibrant A/B are installed.
Purge Calibrant B and recalibrate the
ISE module. If the electrode is new it
may initially drift as it rehydrates over
the course of 15 minutes.