Operator’s Manual: HTI BioChem FC-360
Page 184 of 217
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
cuvette in the optical path and place a white piece of paper against the lens of the photometer.
Adjust the position of the lamp such that the filament can be seen sharply against the paper. In case
of being necessary to fit the focal length moving the lamp, without rotating it (to be able to maintain
the filament in horizontal position), It must look for to focus the beam in the center of the optical path
of the cuvette. Once obtained the adjustment, fix the lamp.
12. Place back the main cover and lids.
13. Run a calibration of the photometer and all the cuvettes
Maintenance
Instrument
Calibration
select
Calibrate photometer
select
Calibrate cuvettes
First cuvette 1
Last cuvette
100
select the position of the solution to use for the calibration ex:
Use container
Calibrate
Figure 15-4
Figure 15-5