Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 75 of 222
Reload Samples
Upon detection of insufficient sample an error notice will appear. The sample process identification is
provided.
Figure 7-50
Figure 7-51
A
No Sample
notice can be seen on the
Operation
Tab, Status field (Figure 7-43).
Figure 7-52
When insufficient sample has been detected, the empty sample tube/cup must be refilled, press
Pause
button, refill sample tube, the patient must be selected on the
Operations Tab
then press the
Reload
Sample
button and press
Pause
button again.
Export Results
Results obtained in the software can be exported to the laboratory information system software through a
standard CSV file (comma, space, value). The file is created according to the process described in Chapter
12, section 12.1.3, and a description of this file data format can be found in Chapter 13, section 13.3
The process can be performed automatically along with the laboratory system software, for further
information contact technical support.