High Technology, Inc.
109 Production Road,
Walpole MA 02081 USA
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared by
J.S. Bolio
Approved by
S. Titov
Effective Date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Page 207 of 222
(0256) Samples and
Reagents Tray error (Home
active)
Sensing failure.
Contact Technical Support.
(0257) Samples and
Reagents Tray error (Home
inactive)
Mechanical failure
Contact Technical Support.
Sensing failure
Contact Technical Support.
FILTER WHEEL
Message ID Number and
Description
Probable cause
Possible Solution
(0302) Filter Wheel not
initialized or initialization error
(Home active)
Home sensing failure.
Contact Technical Support.
Mechanical failure.
Contact Technical Support.
Filter wheel not initialized.
Execute the command to initialize.
(0303) Filter Wheel not
initialized or initialization error
(Home inactive)
Home sensing failure
Contact Technical Support.
Mechanical failure
Contact Technical Support.
Filter wheel not initialized.
Execute the command to initialize.
(0304) Filter Wheel not
initialized
Module not initialized.
Initialize module
(0305) Filter Wheel error
(slotted tray counting error)
Home sensing failure.
Contact Technical Support.
Mechanical failure
Contact Technical Support.
Sensor F out of the counting position.
Contact Technical Support.
(0306) Filter Wheel error
(Home active)
Poor contacts.
Contact Technical Support.
Mechanical failure
Contact Technical Support.
Sensing failure
Contact Technical Support.
(0307) Filter Wheel error
(Home inactive)
Poor contacts
Contact Technical Support.
Mechanical failure
Contact Technical Support.
Sensing failure.
Contact Technical Support.
READING (AMPLIFIERS)
Message ID Number and
Description
Probable cause
Possible Solution