High Technology, Inc.
109 Production Road,
Walpole MA USA 02081
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared By
J.S. Bolio
Approved By
S. Titov
Effective date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
Page 108 of 217
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Saving Controls Files
Controls configuration data in use can be saved from the submenu
Save.
By pressing
Controls Menu
Save
, an exploration window is displayed. Select the folder or location where the data will be saved.
Name the file and
Save
or
Cancel
if the operation will not be carried out.
A file with extension *.aof will be created which can be used as a backup or to be installed in another PC.
Loading Controls Files
Controls configuration data can be obtained through the submenu
Load
from a different file than the file in
use. By pressing Controls
Menu
Load
, an exploration window is displayed. Select the folder or location
from where the data will be entered. Select the file with extension *.aof and press
Open
or
Cancel
if the
operation will not be carried out. All the data of the controls that appear in the selected file will then be
entered.
Warning: By entering configuration data of new controls from the submenu Load, all the data of
controls in use will be deleted.
Before this operation, it is recommended to back up the controls in use as indicated in Chapter 10, section
10.2
11 STATISTICS MENU
The
Statistics Menu
(Figure 11-1), which is accessed from the main screen, is where patients, calibrators
and controls statistical analysis is carried out. There is also a shortcut to the statistical analysis of data
from the
Levy-Jennings Plot
icon.
Figure 11-1
Levy-Jennings of Patients
Levy-Jennings plots can be made to represent values obtained in the same patient for a certain analyte.
To do this, data must be previously ordered in the
PAT Results
Tab, performing an advanced search by
Last Name and Name, so that all processes for the same patient are shown together.
The following steps must be carried out:
1. Select the Process Date from/to which the search will be performed.
2. Perform an advanced search of the data of the same patient (Figure 11-3): For example: Name: