Operator’s Manual: HTI BioChem FC-360
Page 10 of 217
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
If necessary, soak the reference electrode in warm water until the lumen of the electrode has been
cleared of salt build-up.
6. Depress the compression tray to install electrodes in position (as shown in Figure 2-2)
Figure 2-2
7. Press the lock stick and move down the ISE module to put in place
Warning: Do not turn off the analyzer by pressing general button unless maintenance cycle has
been done and electrodes removed.
NOTE: ISE module must be power on all time that the electrodes are in it. Purge cycles are
performed by the ISE module automatically every 30 minutes.
8. Connect tubing to the appropriate containers.